Following our previous GATE-based Monte Carlo simulation studies, a table-top experiment was performed to validate event-by-event image reconstruction for medical imaging. The prototype measured coincidences between 171.3 keV and 245.4 keV non- collinear gamma rays emitted by Indium-111 (111In) point-like source. This imaging modality combines the collimator-based imaging geometry of SPECT with the coincidence detection principle of PET. Based on the simulation results, a tungsten (W) and polylactic acid (PLA) collimator was fabricated using three-dimensional printing and was finally integrated with a small-animal PET system. Imaging was conducted using a 2.0 MBq point- like source (approximately 3.0 mm in diameter) positioned at the center of the field of view. This was followed by scanning multiple point-like sources of varying activities and configurations. Despite the limited spatial resolution of approximately 10.4 mm FWHM transaxially and 15.9 mm FWHM axially, resulting from septal penetration that ultimately broadens the point spread function, the system demonstrated a sensitivity of approximately 1.0 cps/MBq. While the system partially resolved two sources separated by 24.0 mm axially, it failed to resolve separations of 5.0 mm and 10.0 mm transaxially due to collimator material and fabrication limitations. These encouraging results support further optimization and extension of the approach to a full-body PET system.
Recommended Citation
Nkuba, Leonid L.; Olshanoski, Kaylyn; Endo, Satoru; Fukuchi, Tomonori; and Rangacharyulu, Chary
(2026)
"Tomographic medical imaging experiment through coincidence detection of non-collinear cascade gamma-ray emissions from an Indium-111 point-like source,"
Tanzania Journal of Science: Vol. 52:
Iss.
4, Article 4.
Available at:https://doi.org/10.65085/2507-7961.2445
Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 License.